Error Code |
Error description |
Causes and solutions |
1 |
Memory error |
Contact local agent |
No Echo |
Echo loss |
See Action 5 and 6 |
3 |
Hardware error |
Contact local agent |
4 |
Display overflow |
Check settings |
5 |
Sensor error or improper installation/mounting, level in the dead band |
Verify sensor for correct operation and check for correct mounting according to the User’s Manual |
6 |
The measurement is at the reliability threshold |
Better location should be found. |
7 |
No signal received within the measuring range specified in P04 and P05 |
Check programming, also look for installation mistake |
12 |
Linearisation table error: both L(1) and L(2) are zero (no valid data-pairs) |
See ”Linearisation” Section |
13 |
Linearisation table error: same L(i) data is given twice in the table |
See ”Linearisation” Section |
14 |
Linearisation table error: the r(i) values are not monotone increasing |
See ”Linearisation” Section |
15 |
Linearisation table error: measured Level is higher than the last Volume or Flow data-pair |
See ”Linearisation” Section” |
16 |
The check sum of the program is wrong |
Contact local agent |
17 |
Parameter consistency failure |
Check programming |
18 |
Hardware failure |
Contact local agent |